I would guess that many people who read our blog, have worked with medical devices, or have ever had an intravenous injection in a hospital will be familiar with the luer connecter. This small device enables the connection of drug delivery devices and accessories. It might be fair to say that this little invention has improved patient safety like no other.

Not surprisingly when something becomes as popular as the luer connector and usage application grows rapidly, it becomes necessary to control design and specifications. After all the luer connector is intended to standardize connectivity in a large part of the medical device industry. This is achieved by compliance with ISO 80369 Small Bore Connector Standards, of which there are several sub sets that deal with specific types of medical devices including ISO 80369-7 which covers luer connectors and ISO 80369-20 which describes common test methods.

ISO 80369-7 states that:

“Luer connectors shall be evaluated for leakage using either the leakage by pressure decay test method or the positive pressure liquid leakage test method.

“Luer connectors evaluated for fluid leakage by pressure decay test method shall not leak by more than 0.005 Pa*m3/sec while being subject to an applied pressure of between 300 kPa and 330 kPa over a hold period between 15s and 20s using air as the medium. Manufacturers may use a greater pressure.”

Compliance may be checked by using the tests specified in Annex B of ISO 80369-20. These require the reporting of the pressure change and the leak rate.

Recently we were asked by a client to supply a leak test instrument that would facilitate compliance with part 20 and that could automatically calculate many of the measurements that can be time consuming and vulnerable to human error.

Some of the other specific requests included features such as:

  • Pressure and vacuum decay measurements
  • Automatically calculate the test part and test circuit volume in m3
  • After completing the pressure change measurement and volume calculations automatically calculate the test part leak rate in Pa*m 3/sec
  • Then compare the calculated leak rate to the specification’s maximum allowable and provide a clear Accept or Reject decision to the operator
  • Use engineering units as called out in the specification
  • Follow specification as close as possible to eliminate the need to request a process deviation
  • And let’s not forget they requested the data for all steps in the process be stored to their server

These are the laws of physics that were used to satisfy our customer’s requests:

Ideal Gas Law Review

•             Leak testing techniques are strongly influenced by the physical properties of gases

•             The behavior of a gas can be expressed with the basic Ideal Gas Law (Emile Clapeyron in 1834) PV = nRT. The state of an amount of gas, n, is determined by pressure, volume and temperature.

Boyles Law Review

•             Leak Rate Formula can then be tweaked to solve for volume if Leak Rate and Pressure Change are known:

For Example:

                Leak Rate            3 cc/min

                Time                      10 seconds (.16667 Minutes)

                Atmos. press.    1013.2 mbar

                Press. Change   20 mbar

                Total Volume   =25.33 cc

The OptimavT was chosen for this application because of its powerful PROGRAM JUMP, MATH and STEP functions.  By working closely with the client’s quality assurance and regulatory affairs teams our application development team developed a series of program steps and register entries that would perform the necessary calculations automatically and, using the clients own accept reject criteria, segregate good and bad parts by pass and reject results.

To solve the leak rate equations and make the Accept/Reject decision the following step types were utilized:

OptimavT Measure Pressure Change- Calculating Volume and Pressure Leak Rate

OptimavT MATH- Calculating Volume

OptimavT Measure Vacuum Change- Calculating Vacuum Leak Rate

OptimavT Math- Calculating Leak Rate

OptimavT MATH- Engineering Unit Conversions

Here is a snippet of the OptimavT register Set 1 showing some of the registers used to perform these tasks:

NB: the above snippet is only a few lines of register entries and calculations.

It is anticipated this will save the customer hundreds if not thousands of man hours in the lab by automating the compliance check of their products to this specification. How many hours might you save with a setup such as this?  

This application is an excellent example of the thinking behind the design of the OptimavT. Our product development team designed an instrument so versatile that it can combine the attributes of a leak tester, a PLC and a computer in one neat package.

 For aiding compliance, taking the headache out of leak test calculations, removing the scope for human error, and providing fast, reliable results to complex problems, OptimavT leads the class.

If you have a complex leak or flow testing application that you use some help with please contact me at 281-671-2000 or email me at joe.pustka@uson.com


Joe Pustka

Director Applications Engineering

Contact Us

Certifications


Uson L.P. - Corporate

8640 N. Eldridge Parkway
Houston, TX 77041 USA

Tel: +1-281-671-2000
Fax: +1-281-671-2001

Email: info@uson.com

Uson UK – Europe

Western Way
Bury St. Edmunds
Suffolk, IP33 3SP United Kingdom

Tel: +44-1284-760606
Fax: +44-1284-763049

Email: salesuk@uson.com

Uson – Detroit

31471 Utica Rd
Fraser, MI 48026 USA

Tel: +1-248-589-3100
Fax: +1-248-589-3103

Email: Michigan@uson.com

Uson – China

Room 601, 567 Lan Gao Road
Shanghai, 200333, China

Main: 021-583 65859, 021-583 59187
Fax: 021-583 59185, 021-337 73559

Email: sales@uson.com.cn